Optimizing patient safety: an evaluation of computed tomography dose-length product variations across hospitals in Abuja, Nigeria

Authors

  • Joseph Teryima Iortile Department of Physics, Faculty of Sciences, National Open University of Nigeria, Abuja, Nigeria
  • John Kahugu Simon Department of Physics, Faculty of Sciences, National Open University of Nigeria, Abuja, Nigeria

Keywords:

Computed tomography, Dose-length product, Diagnostic reference levels, Radiation optimization, Abuja

Abstract

The increasing use of computed tomography (CT) as a diagnostic tool in Nigeria raises concerns about high radiation doses and potential cancer risk. To address this concern, CT dose-length product (DLP) was evaluated at ten hospitals in Abuja, Nigeria, using polymethyl methacrylate phantoms, a 100 mm pencil ionization chamber, an electrometer, and a CT probe. The study assessed variations in radiation dose across the facilities by measuring DLP in mGy.cm for chest and head phantoms. Chest phantom DLP values ranged from 79.36 ± 0.001 mGy.cm at Hospital H to 677.40 ± 0.01 mGy.cm at Hospital D. Head phantom values ranged from 230.14 ± 0.004 mGy.cm at Hospital A to 1165.73 ± 0.3 mGy.cm at Hospital I. Most facilities recorded DLP values below the European diagnostic reference levels (DRLs) used for comparison. However, Hospital D exceeded the chest reference level of 650 mGy.cm, while Hospitals E and I exceeded the head reference level of 1050 mGy.cm. These findings indicate substantial variation in scanning protocols among the facilities. The study recommends routine quality control, equipment calibration, standardized protocols, and refresher training for CT operators to ensure doses administered to patients are As Low as Reasonably Achievable (ALARA) principle and reduce unnecessary radiation exposure.

22 18

References

[1] H. Mohammad, J. T. Iortile, I. Garba & M. A Suwaid. ``Knowledge of radiation and its effects among Doctors in Makurdi, North Central Nigeria'', International Research Journal of Basic and Clinical Studies, 1(7) (2013) 103. http://www.interesjournals.org/IRJBCS.

[2] H. Mohammad, J. T. Iortile, T. T. Bitto, E. O. Eje-Ejum & B. I. Jayeoba. ``Social-Demographic Characteristics and Awareness of Radiation Hazards among Radiology Technicians and Sub staff of Radiology Units in Benue State'', Nig Hosp Pract, 14 2014 28. http://www.ajol.info/index.php/nhp.

[3] J. T. Iortile, L. Alumuku & I. D. Shagu, ``Comparative review of conventional computed tomography and photon counting detector-based computed tomography systems'', FUAM Journal of Pure and Applied Science 5 (2025) 82.

[4] L. Alumuku, J. T. Iortile & T. Daniel, ``Analysing spectral imaging at different photon energy bins for breast lesion contrast visualisation from computed tomography images'', Nigerian Journal of Physics 33 (2024) 127. https://doi.org/10.62292/njp.v33i4.2024.321. DOI: https://doi.org/10.62292/njp.v33i4.2024.321

[5] J. T. Iortile, L. Alumuku & A. Shaanika, ``Radiation dose optimization in computed tomography examinations: A comparative analysis of body and head phantoms in some radiological units of Abuja hospitals'', FUAM Journal of Pure and Applied Science 5 (2025) 36. Available online: https://www.researchgate.net/publication/390607605_Radiation_Dose_Optimization_in_Computed_Tomography_Examinations_A_comparative_Analysis_of_Body_and_Head_phantoms_in_some_Radiological_Units_of_Abuja_Hospitals.

[6] American Association of Physicists in Medicine (AAPM), ``AAPM Computed Tomography Radiation Dose Education Slides'' (2013). http://www.aapm.org/pubs/CTprotocol.

[7] T. A. Mahmood & S. R. Hassan, ``Optimization of CT radiation dose: Insight into DLP and CTDIvol as scanner output metrics'', Future Health Journal, 11(3) (2024) 204. https://futurehealthjournal.com/optimization-of-ct-radiation-dose-insight-into-dlp-and-ctdi/.

[8] J. T. Iortile & T. A. Ige, ``Measurement of computed tomography dose quantities at some radiological units of Abuja hospitals'', African Journal of Medical Physics 4 (2022) 48. https://www.researchgate.net/publication/376720806_Number_1_48_FAMPO_Federation_of.

[9] B. Maidamma, I. Maina, O. W. Olasoji, A. M. Tijjani, B. F. Namq, O. A. Adeyeba, M. Mujahid, U. Rilwan, M. I. Sayyed & T. Y. Wais, ``Computed tomography procedures and current dose practice determination in Sokoto State CT centres'', Research Square preprint (2025). https://doi.org/10.21203/rs.3.rs-7273866/v1. DOI: https://doi.org/10.21203/rs.3.rs-7273866/v1

[10] J. T. Iortile & L. Alumuku, ``The significance of medical physicists in ensuring regulatory adherence and facility architecture in medical diagnostics'', Savanna Journal of Basic and Applied Sciences 7 (2025) 202.

[11] J. D. Zira, C. C. Nzotta & J. D. Skam, ``Diagnostic reference levels (DRLs) for computed tomography (CT) examinations in north eastern Nigeria'', Pakistan Journal of Radiology 27 (2017) 364. https://pakjr.com/index.php/PJR/article/view/750.

[12] American College of Radiology, ``Radiation dosimetry: CT (revised 4 June 2025)'', ACR Accreditation Support (2025). https://accreditationsupport.acr.org/support/solutions/articles/11000056198-radiation-dosimetry-ct.

[13] E. Samei, D. Bakalyar, K. L. Boedeker, S. Brady, J. Fan, S. Leng, K. J. Myers, L. Popescu, J. C. Ramirez Giraldo, F. Ranallo, J. Solomon, J. Vaishnav & J. Wang, ``Performance evaluation of computed tomography systems: Summary of AAPM Task Group 233'', Medical Physics 46 (2019) e735. https://doi.org/10.1002/mp.13763. DOI: https://doi.org/10.1002/mp.13763

[14] E. U. Ekpo, T. Adejoh, J. D. Akwo, C. E. Owujekwe, A. M. Ali, M. Abba, K. A. Adesina, D. O. Omiyi & U. H. Chiegwu, ``Diagnostic reference levels for common computed tomography (CT) examinations: Results from the first Nigerian nationwide dose survey'', Journal of Radiological Protection 38 (2018) 525. https://doi.org/10.1088/1361-6498/aaaaf8. DOI: https://doi.org/10.1088/1361-6498/aaaaf8

[15] A. Ajetunmobi, T. W. David, A. T. Talabi, S. K. Alausa, A. P. Ejiwunmi, O. J. Odogboro & E. B. Peter, ``Estimation of effective dose for patients undergoing computed tomography scans from selected diagnostic centers in southwestern Nigeria'', FULafia Journal of Science and Technology 10 (2026) 197. https://doi.org/10.62050/fjst2026.v10n2.384. DOI: https://doi.org/10.62050/fjst2026.v10n2.384

[16] W. Igoniye, M. A. Briggs-Kamara, I. Cookey, F. B. Sigalo & C. F. Njeh, ``Establishing local diagnostic reference levels for computed tomography in a Nigerian metropolis: A retrospective analysis of clinical practice and radiation dose optimization'', Egyptian Journal of Basic and Applied Sciences 12 (2025) 567. https://doi.org/10.1080/2314808X.2026.2624216. DOI: https://doi.org/10.1080/2314808X.2026.2624216

[17] I. Yemisi, T. H. Okafor, I. Pada, O. O. Samson, A. Abdul-Jamiu & S. H. Sarki, ``Optimization of patient radiation protection and dose levels in some Nigerian CT facilities using OSL dosimetry'', American Journal of Physics and Applications 14 (2026) 32. https://doi.org/10.11648/j.ajpa.20261403.11. DOI: https://doi.org/10.11648/j.ajpa.20261403.11

[18] M. Arfat, A. Haq, T. Beg & G. Jaleel, ``Optimization of CT radiation dose: Insight into DLP and CTDI'', Future Health 2 (2024) 148. https://doi.org/10.25259/FH_45_2024. DOI: https://doi.org/10.25259/FH_45_2024

FIG3

Published

2026-11-02

How to Cite

Optimizing patient safety: an evaluation of computed tomography dose-length product variations across hospitals in Abuja, Nigeria. (2026). Recent Advances in Natural Sciences, 4(2), 415. https://doi.org/10.61298/rans.2026.4.2.415

How to Cite

Optimizing patient safety: an evaluation of computed tomography dose-length product variations across hospitals in Abuja, Nigeria. (2026). Recent Advances in Natural Sciences, 4(2), 415. https://doi.org/10.61298/rans.2026.4.2.415